Coa LAB 6000, Fully automated coagulation analyzer for all clotting assays
A simple batch and random analyzer for clotting tests

Specifications		: 	
Measuring principle 		:	The opto-mechanical endpoint method features turbidimetry 
				allowing the use of lipemic and/or coloured samples with kaolin.
Measuring station		:	6 measuring channels, primary rotor for 6 cuvette cartridges, each 
				cartridge includes 6 cuvettes, 18 sample positions, 6 reagent positions.
Test volume		:	125 µl -250 µl
Traverse			:	Movement into 2 directions X-Z for pipetting and transport of plasma and reagent. 
				Pipettor heated to 37 °C.  A level sensor detects the presence of liquid.
Dilutor			:	250 ml syringe, 1 pipetting tube
Liquid positions		:	6 positions for reagent or deficiency plasmas and washer at room 
				temperature. All reagent positions can be stirred.
Software			:	Method menu for PT, aPTT and Fibrinogen. Documentation of all test parameters
				Calibration curves, date, time, conversions of %, g/l, mg/dl, ration, INR
				and incubation time can automatically be loaded by ChipCARD.
				External service program.
Display			:	Alphanumeric LCD-Display, 2 lines, 20 characters each.
Operating modes		:	Batch-Mode for PT/aPTT or Fibrinogen. Combi-Mode incl. PT/aPTT and 
				Fibrinogen in one cuvette cartridge.
Throughput		:	>120 PT/h
Special features		:
				* Single and double detection adjustable, Combi-Mode allows 
       				* PT/aPTT and Fibrinogen measurements at one time.
				* On board Waste Box for used cartridges.
				* Plasma or reagent start ( can be configured )
				* Fibrinogen ( Claus method )
				* Optimised cuvette cartridge consumption
Printout			:	Built-in thermal printer.
Barcode reader		:	For fast introduction of patient/sample numbers.
Interfaces			:	1x  RS232 C Host/Service
				1x  Barcode Scanner
Power requirements		:	230V / 115V, 50-60 Hz, 150 VA
Dimensions		:	72 x 45 x 55 cm
Weight			:	± 17 kg
Disposable			:	Cartridges, incl. mixers, reagent containers, plasma cups.