Coa LAB 6000, Fully automated coagulation analyzer for all clotting assays Specifications : Measuring principle : The opto-mechanical endpoint method features turbidimetry allowing the use of lipemic and/or coloured samples with kaolin. Measuring station : 6 measuring channels, primary rotor for 6 cuvette cartridges, each cartridge includes 6 cuvettes, 18 sample positions, 6 reagent positions. Test volume : 125 µl -250 µl Traverse : Movement into 2 directions X-Z for pipetting and transport of plasma and reagent. Pipettor heated to 37 °C. A level sensor detects the presence of liquid. Dilutor : 250 ml syringe, 1 pipetting tube Liquid positions : 6 positions for reagent or deficiency plasmas and washer at room temperature. All reagent positions can be stirred. Software : Method menu for PT, aPTT and Fibrinogen. Documentation of all test parameters Calibration curves, date, time, conversions of %, g/l, mg/dl, ration, INR and incubation time can automatically be loaded by ChipCARD. External service program. Display : Alphanumeric LCD-Display, 2 lines, 20 characters each. Operating modes : Batch-Mode for PT/aPTT or Fibrinogen. Combi-Mode incl. PT/aPTT and Fibrinogen in one cuvette cartridge. Throughput : >120 PT/h Special features : * Single and double detection adjustable, Combi-Mode allows * PT/aPTT and Fibrinogen measurements at one time. * On board Waste Box for used cartridges. * Plasma or reagent start ( can be configured ) * Fibrinogen ( Claus method ) * Optimised cuvette cartridge consumption Printout : Built-in thermal printer. Barcode reader : For fast introduction of patient/sample numbers. Interfaces : 1x RS232 C Host/Service 1x Barcode Scanner Power requirements : 230V / 115V, 50-60 Hz, 150 VA Dimensions : 72 x 45 x 55 cm Weight : ± 17 kg Disposable : Cartridges, incl. mixers, reagent containers, plasma cups.
|