The D.C. Analyzer named hereafter as analyser, is a fully automated photo optical blood plasma hemostasis instrument capable of performing clotting, chromogenic* and immunoassays* simultaneously in random access or batch mode from primary and secondary tubes.
For in-vitro diagnostic use only!
* Depending on the analyser model the system is capable to perform a wide range of coagulometric, chromogenic and immunologic coagulation tests such as Prothrombin time, activated and partial Thromboplastin time, Fibrinogen, derived Fibrinogen and special tests such as single factor assays, Anti-Thrombin III, Protein C, Protein S, C-Reactive Protein, DDimer and others based on the wavelength available in different analyser types. However; the manual will indicate any difference between the available model types if required.
The analyser is intended for use by trained operators in coagulation labs for in-vitro diagnostics.
The analyser represents a new the state of the art development and is a microprocessor-controlled, two-channel photo-optical fully automated blood plasma hemostasis analyser. The analyser is a modular construction.

- 1 - Pipettor
- 2 - Cup Array for Sample Cups (Controls/Calibrators)
- 3 - Reagent Block (removable)
- 4 - Washer / Cleaner Position
- 5 - Rinsing Station
- 6 - Measuring / Cuvette Ring Rotor
- 7 - STAT Positions
- 8 - Sample Positions
- 9 - USB Stick Interface (underneath) + handles to carry analyzer
- 10 - Touch sensitive colour display
- 11 - Safety Shield Handling
Technical Data
General Description
- Type of instrument Fully automated coagulation analyzer
- Application Coagulation, Chromogenic, Immunologic tests
- Working methods Sample oriented Test selective
- STAT (emergency)
- Measuring principle Photometric; with magnetic stirrers to homogenize the test mixture and increase the sensitivity
- Light source LED, light emitting diode
- Cuvette Volume 150μl - max. 300μl
- Test throughput PT 120/h, aPTT 60/h, +/- 20 tests/h
- Cuvette volume min. 150 μl, max. 300 μl (test suspension)
- Measuring channels Two measuring channels at - one wavelength 405nm or - three wavelength 405/570/740nm
- Calibration - Manual input of calibration points (max. 7 points),
- two calibration curves for each test or Automatic calibration
- STATS STAT loading is possible at any time in up to three assigned positions. Routine testing can be interrupted for STAT loading and then automatically giving priority to the STAT sample
Analyzer data
- Reagent positions 15 positions at room temperature
- - (R1-R3) for 15ml reagent vials (left/right pos. stirred)
- - (R4-R15) for 5ml reagent vials.
- Reagent block can be removed from working area for external cooling.
- Washer positions 2 positions at room temperature
- - (W1-W2) for Washer & Cleaner
- Calibration/Dilution 8 positions at room temperature
- - (C1-C8) for Controls, Calibrators & Dilutions
- Sample positions 2x11 positions for primary & secondary tubes/cups
- 1x3 positions for STAT
- Measuring rotor One heated measuring rotor at 37°C +/-0.4°C with two measuring channels, designed for cuvette rings
- Cuvette rings Cuvette ring with 32 cuvettes
- Wash Station One wash station for pipettor cleaning
- Pipettor Capacitive, movable in three directions (x, y, z),
- heated at 37°C
- Dilutor 250μl syringe
- - minimal dosing 2μl
- - maximal dosing 250μl
- - Resolution: 12 half steps for 1 μl (for 250 μl syringe)
- - Level sensors for distilled and waste water container
- Rinsing Two modes
- - Regular with distilled water
- - 4 intensive with washing solution for probe cleaning
- (Thrombin inactivator)
- Cleaning Cleaning of pumps and tubing with Cleaner
- Display Touch sensitive colour display
- Visible area: 11.5x8.7cm
- Display resolution: 320 x 240 pixels, ¼ VGA).
- Software Linux based operating system with Graphic User Interface (GUI)
- Programmable 50 reagents/tests can be freely programmed
- Methods
- Interfaces 1x USB for external printer device
- 1x USB for USB Memory device
- 1x LAN for external PC connection
- 2x RS 232C serial interfaces
- 1x external Barcode Scanner connection
Physical data
- Voltage: 100 - 240 V
- Frequency: 50 - 60 Hz
- Capacity: 300 VA
- Fuses: 5A/250 V time lag (T5A L250V)
- IEC 60127-2/III
- Noise emission < 56 dB (A)
Ambient Conditions
- Usage: Inner rooms only, dry environment
- Altitude: ≤ 2.000 meters
- Mains voltage fluctuation: Max. ± 10% of rated voltage
- IP Code (IP20):
- accord. IEC 60529
- 2 = < 12.5mm, Fingers or similar objects
- 0 = Not protected against harmful ingress of water
- Pollution degree: Class 1
- Temperature: during operation +15°C - +30°C
- during transport + 2°C - +50°C
- Humidity: 10% - < 85%, non-condensing
Dimensions
- Instrument (WxHxD): 78 x 58 x 50 cm 30.7x 22.8 x 19.7 in.
Shipping:
Weight
- Instrument: 25 kg 55 lb
- Shipping: 30 kg 66 lb